L’unità PTB330S combina un barometro PTB330 con un indicatore portatile MI70 (che alimenta gli strumenti e fornisce il display per visualizzare i dati) ed (opzionale) una sonda HMP155 per temperatura ed umidità. Lo strumento è adatto ad effettuare misure di riferimento in ambito meteorologico e industriale; viene dotato di valigetta da trasporto weatherproof per una facile movimentazione.

Specifiche Tecniche

Barometric Pressure range500…1100 hPa
Barometric Pressure linearity±0.05 hPa
Barometric Pressure hysteresis±0.03 hPa
Barometric Pressure repeatability±0.03 hPa
Barometric Pressure calibration uncertainty±0.07 hPa
Barometric Pressure accuracty at +20 °C±0.10 hPa
Barometric Pressure temperature dependence±0.01 hPa
Barometric Pressure total accuracy at -40…+60 °C±0.15 hPa
Barometric Pressure long-term stability (/year)±0.1 hPa
Barometric Pressure response time2 s
Humidity measurement range0…100% RH
Humidity accuracy at +15…+25 °C, 0…90% RH±1% RH
Humidity accuracy at +15…+25 °C, 90…100% RH±1.7% RH
Humidity accuracy at -10…+40 °C±(1.0 + 0.008 x reading)% RH
Humidity Factory Calibration Uncertainty (+20 °C), 0…40% RH±0.6% RH
Humidity Factory Calibration Uncertainty (+20 °C), 40…97% RH±1.0% RH
Humidity response time (63%) at +20 °C in still air, with a sintered PTFE filter20 s
Humidity response time (90%) at +20 °C in still air, with a sintered PTFE filter60 s
Temperature measurement rage-10…+40 °C
Temperature accuracy at -10…+20 °C±(0.176 – 0.0028 x temperature) °C
Temperature accuracy at +20…+40 °C±(0.07 + 0.0025 x temperature ) °C
Temperature response time (63%) with additional temperature probe in 3 m/s air flow< 20 s
Temperature response time (90%) with additional temperature probe in 3 m/s air flow< 35 s
Power supplyRechargeable NiMH battery pack with AC-adapter or 4xAA-size alkalines, type IEC LR6
MI70 probe ports2
MI70 data interfaceRS-232
PTB330 supply voltage10…35 VDC (if not powered by MI70)
PTB330 data interfaceRS-232C
HMP155 data interfaceRS-485
Operating temperature range-10…+40 °C

Per maggiori dettagli sul prodotto, download aggiuntivi, FAQ, accessori ed esempi di applicazioni visitare il sito del produttore alla pagina PTB330S